BIOPHYSICS
Nanometer-localized multiple single-molecule fluorescence microscopy

,

,
, ||
Departments of *Physics, ¶Molecular Genetics and Cell Biology, and
Chemistry and
Institute for Biophysical Dynamics, University of Chicago, Chicago, IL 60637
Edited by Robert Haselkorn, University of Chicago, Chicago, IL, and approved June 23, 2004 (received for review March 26, 2004)
| Abstract |
|---|
|
|
|---|
![]() | [1] |
where
is the wavelength of the collected photons and N.A. is the numerical aperture of the system. Using visible light (i.e.,
500 nm) and a high N.A. objective (i.e. >1) allows achieving
250-nm lateral resolution. However, most biological macromolecules (DNA, RNA, proteins, etc.) and molecular machines (e.g., ribosome and spliceosome) are much smaller in two or all three dimensions, with important features or subassemblies lying within 10 nm of each other (2).
Fluorescence resonance energy transfer (FRET) is widely used for studying biological systems on the nanometer scale (i.e., <10 nm; ref. 3). FRET occurs over distances similar to the Förster radius, which is
5 nm for common fluorophores (4). However, FRET only gives the relative distance between two probes, whereas their absolute positions remain unknown. Centroid localization, which has been used for single-particle tracking, allows determination of the centroid position of the particle to a much better precision than the length scale defined by the Rayleigh criterion (57). Recently, this approach to localizing fluorophores has been applied to single dye molecules (Cy3 and Rhodamine dyes) bound to molecular motors, and 2-nm precision in localization has been achieved when an oxygen scavenging buffer was used to suppress photobleaching (8, 9). However, centroid localization can only be used for isolated single molecules and it does not improve the resolution of the imaging system. It has been shown that two probes with different fluorescent wavelengths can be resolved within 10 nm because they are spectrally distinguishable (10). This method does not help when two or more objects with the same emission spectrum are within a diffraction-limited spot. Imaging applications beyond even the capabilities of "superresolution" microscopy (11) and without the complexity inherent to near-field scanning optical microscopy (12), such as the distribution and disposition of components of molecular complexes and "machines" (2) and their motions, would benefit from approaches with few nanometer resolution.
Whereas DNA sequencing provides detailed information that is essential for understanding the function and evolution of genes and their flanking regions, whole-genome characterization provides the genetic context in which these functions and changes take place (1315). Although much progress has been made in the assembly of large-genome regions from raw sequence data, the short scope of individual sequencing reactions (on the order of 1 kb) makes complete genome characterization by sequencing alone an extremely challenging task (16, 17). Optical mapping of much longer (1001,000 kb) single DNA molecules, by using gaps generated by digestion with restriction endonucleases as landmark sites, has provided a breakthrough method of comparatively rapid whole-genome characterization with the potential to complement and support whole-genome sequencing efforts (18). Whereas optical mapping provides a valuable approach for comparative genomics, the information content of these maps is restricted by dependence on an imaging method whose resolution is limited by the Rayleigh criterion. Given an imaging method with much better resolution, optical mapping of DNA molecules by using fluorescent tags bound to specific sequence-recognition sites could be exploited to dramatically increase the information content of long-range physical mapping and of whole-genome characterization.
This paper demonstrates nanometer-localized multiple single-molecule (NALMS) fluorescence microscopy, a method that makes use of centroid localization and exploits the inevitable photobleaching of single fluorophores to achieve nanometer-scale resolution of single-molecule fluorophores. The resolution obtainable with this approach is demonstrated for two identical fluorophores (i.e., Cy3) separated by 8 nm with 2.5-nm precision for the distance determination and depends on the total number of photons collected for each fluorophore. The method can be easily extended to multicolor applications. This paper also demonstrates that NALMS fluorescence microscopy is an attractive approach for ultraresolution DNA mapping. Single fluorophores are attached to specific 7-base sequences on DNA by peptide nucleic acid (PNA) labeling and NALMS fluorescence microscopy is used to determine the fluorophore locations with <10-nm (i.e., 30-base) resolution.
| Materials and Methods |
|---|
|
|
|---|
DNA labeling and surface immobilization was performed with some modification of published procedures (20, 21). No. 1 coverslips were soaked in 2% hydrofluoric acid solution for 10 min, rinsed with water, and dried with nitrogen gas. A small chamber was made with two coverslips and 3M double-sided tape. Fifty microliters of 1 mg/ml BSA-biotin (Sigma, catalog no. A-8549) in TE buffer was added to the chamber, allowed to sit for 10 min, and was then washed with 100 µl of TE buffer. Fifty microliters of 0.2 mg/ml streptavidin (Molecular Probes, catalog no. S-888) in TE buffer was then added, allowed to sit for 10 min, and the protein-coated coverslip was washed with 100 µl of TE buffer. A 3 nM solution of Cy3- and biotin-labeled 24-base or 40-base duplex DNA solution in TE buffer was mixed with 10 mg/ml BSA (Behring Diagnostics, catalog no. 126575) by 1:9 (vol/vol). Fifty microliters of the mixture was added to the chamber, allowed to sit for 10 min, and was then washed with 200 µl of TE buffer. One hundred microliters of imaging buffer was added and the chamber was sealed with nail polish. The imaging buffer contains TE buffer with 30 mg/ml glucose, 0.1 mg/ml glucose oxidase (Sigma-Aldrich, catalog no. G-7016), 0.2 mg/ml catalase (Roche Diagnostics, catalog no. 106810), 1% (vol/vol) 2-mercaptoethanol, and 10 mM MgCl2. These enzymes remove molecular oxygen in solution, thereby reducing the rate of irreversible photobleaching of the dye molecules (8).
Sample Preparation for DNA Mapping. Coverslips were cleaned as described for the DNA ruler measurement. Fifty microliters of 0.015% 3-aminopropyltrimethoxysilane (Aldrich, catalog no. 28177-8) in acetone was sandwiched between two coverslips and incubated for 15 min. The coverslips were separated, rinsed with acetone, then with water, and were then dried with nitrogen gas. A drop of Alexa 488-bis-PNA-labeled
-DNA solution was placed on one functionalized coverslip. The DNA was spread by dynamic molecular combing by rolling the drop across the coverslip (22). Fifty microliters of 20% 2-mercaptoethanol was added and covered with another coverslip. This sandwich was sealed with nail polish.
Binding fluorophore-labeled bis-PNA to DNA is accomplished by incubating the mixture of 12 µM Alexa 488-bis-PNA (sequence: Alexa 488-eg1-TTTCTCC-eg1eg1eg1-CCTCTTT-amide; ref. 23) and 12.5 nM
-DNA (New England Biolabs, catalog no. N3013L) in 100 mM phosphate buffer (30 mM Na2HPO4/70 mM NaH2PO4, pH 6.6) at a decreasing temperature, beginning at 80°C and decreasing 1°C every 5 min to 4°C. The DNA/PNA complexes were separated from free PNA by ethanol precipitation and were then resuspended in water.
Total Internal Reflection Fluorescence (TIRF) Microscope. All single-molecule- and dual-fluorophore-labeled DNA ruler test data were obtained by using a home-built objective-type TIRF microscope, based on an Olympus (Melville, NY) IX71 inverted fluorescence microscope and a x60 objective (Olympus, N.A. = 1.45, oil immersion). A 532-nm laser [New Focus (San Jose, CA), model no. 3951-20, 20 mW power] was used as the illumination source. An OD = 1 neutral density filter was used for most experimental results shown. The incident laser power at the TIR interface is roughly 0.5 mW, illuminating a circular region 40 µm in diameter. Due to the Gaussian profile of the laser beam, the illumination intensity is higher in the center of the area than at the periphery. Thus, the emission intensity (number of photons collected per unit time) depends on the location of dye molecules. Furthermore, even within a diffraction-limited spot, the fluorescence emission intensity of individual dye molecules may exhibit variability, presumably resulting from differences in the local environment and the way the dyes bind to the DNA and orientations created by the BSA/streptavidin/biotin double tether. The emitted photons were collected through a filter cube [Chroma (Rockingham, VT) Q560LPBS, HQ585/40M], magnified by a x3.3 eyepiece (Olympus) and detected by a back-illuminated, TE-cooled, frame-transfer charge-coupled device (CCD) detector [Andor, DV435-BV (South Windsor, CI)] with 1,024 x 1,024 13-µm square pixels. The effective pixel size of the CCD detector due to magnification of the x60 objective and the x3.3 eyepiece is 65.7 nm (which is parameter a in Eq. 3). The width of the point spread function of the setup (Sx and Sy in Eq. 2) is Sx,y = 1.9 pixels, which is 124.8 nm. One count from the CCD detector corresponds to roughly two collected photons (Andor system performance booklet). For most of the data shown, the CCD was operated in frame transfer mode, cooled to 50°C, with 1 s of integration time for each frame and <1 ms of dead time between frames (i.e., the frame transfer time).
Centroid Localization. It has been shown that the centroid of a diffraction-limited image of a point object can be localized to high precision by fitting the collected distribution of photons/counts (i.e., the image) to the point spread function of the system (24). This is usually well approximated by a two-dimensional Gaussian fit,
![]() | [2] |
where (x0, y0) is the centroid of the point source, Nxy and N00 are the counts (or photons collected) at pixel (x, y), and at the centroid position (x0, y0) on the CCD detector, Sx and Sy are the width of the point spread function in the x and y directions, and B is the baseline. The baseline includes the CCD detector background, readout noise, and residual scattered light. The SE in centroid localization is well estimated by
![]() | [3] |
where i = x or y, N is the total counts for the image, a is the effective pixel size of the detector, and b is the SD of background noise.
Later in this paper, we will study the distance between two fluorophores. Because distance is calculated from the centroid location of two fluorophores as
![]() |
the SE in the distance measurement can be calculated from the uncertainty of the centroid localization of each of the two fluorophores (Eq. 3) as in ref. 8:
![]() | [4] |
The 1-s integration time used in the present experiment is much longer than the period of the thermal motion of the fluorophores, DNA and BSA/streptavidin/biotin tether. Assuming the thermal motion is symmetric about the tether point, it will not affect the centroid determination of the dye position or the distance measurement between two dyes.
Photobleaching of Single Fluorophores. Single fluorophores have the signature of almost constant emission intensity when the collection time is much longer than the "blinking" period and terminating with abrupt photobleaching (25). Photobleaching is a Poisson process that obeys an exponential distribution,
![]() |
where P(t) is the probability that the fluorophore photobleaches at time t, while
is the time constant for photobleaching. Note that
depends on many molecular photophysical parameters (absorption cross section, intersystem crossing time, etc.) and external variables (e.g., focused laser intensity, concentration of singlet oxygen, etc.). Thus, for a collection of a few fluorophores within a diffraction-limited spot, sudden step changes in fluorescence intensity will be observed with the number of steps equal to the number of fluorophores.
| Results and Discussion |
|---|
|
|
|---|
|
|
|
Verification of Nanometer Localization with Nanoscale Rulers. To test the accuracy of NALMS microscopy, 24- and 40-base double helix DNA molecules labeled at both ends with Cy3 dyes and biotin are used as nanoscale rulers (see Materials and Methods). A series of images are taken until all of the fluorophores photobleach. NALMS position analysis is used to resolve the multiple dyes as described in the previous subsection and the distance between the dyes is calculated.
Fig. 4 is the histogram for the end-to-end distance of a 24-base DNA ruler measured by NALMS fluorescence microscopy. Because the distance cannot have negative values and the mean distance is very close to the boundary value (i.e., zero), the distance distribution may exhibit a skew. As a result, a lognormal distribution may be a more appropriate description of the histogram than a Gaussian distribution (26). The lognormal fit to the histogram gives the end-to-end distance as [82, 8 + 3] nm where 8 nm is the center of the lognormal fit. The 2- and 3-nm values are the width of the fitted lognormal function for the region to the left and to the right of the center value, respectively, and are a measure of the SE of the distance measurement. The expected end-to-end distance of a 24-base double-stranded B-form DNA ruler is 8 nm [24 bases x (1 nm/3 bases)]. Therefore, the measured result agrees very well with the expected value.
|
The present experiments employ a CCD detector with effective pixel size of a = 65.7 nm and Sx,y = 124.8 nm. The events in Fig. 4 only include those dual-fluorophore traces with >500 accumulated counts at the peak pixel for the part of the trace used for centroid localization. The precision for centroid localization is better for the events with more counts collected, and thus the SE of the distance measurement in the histogram is mostly defined by the events close to the chosen threshold of collected counts. The threshold value corresponds to total counts N = 11,300 and the SD of the background is b = 13 counts. According to Eq. 3, this corresponds to
x0 =
y0 = 1.8 nm for each fluorophore. Thus, Eq. 4 gives the SE of the end-to-end distance measurement as 
d2
1/2 = 2.5 nm, which agrees well with the measured SE, i.e., the width of the lognormal fit of Fig. 4.
The above analysis assumes that the behavior of the fluorophores is time-invariant. If this assumption fails, it would be a source of error in position determination that must be assessed. As seen from Figs. 2 and 3, the fluorophore mean intensity is generally constant with statistical fluctuations. The variation in intensity of fluorophore 1 does not affect the localization result of its own centroid and only affects the fitted value of N00. Any change in the emission intensity or position of fluorophore 1 between intervals 1 and 2 of the time trace would affect the calculated centroid location of fluorophore 2. A larger difference in the average emission behavior of fluorophore 1 would cause a larger error in the localization of fluorophore 2. This effect is illustrated in Fig. 5. Beginning with an individual distance measurement in the data for Fig. 4, in which each fluorophore had 500 counts for the peak pixel (measured distance in this instance was 5.9 ± 2.5 nm), the estimated intensity of fluorophore 1 was varied before subtraction from the data used to locate fluorophore 2. As seen from Fig. 5, the deviation of the distance estimate is within 1 nm for <10% variation in the intensity of fluorophore 1. For most events included in Fig. 4, the deviation of each image from its average behavior is <12%, and usually the average behavior of fluorophore 1 is taken as the average of five or more images, and thus the average obtained this way has <12%·(5)1/2 = 5% SE from the true average value. According to Fig. 5, this result corresponds to <0.3-nm deviation in the centroid localization of fluorophore 2. This deviation is small compared with the other errors estimated from Eq. 4. Thus, Eq. 4 is a good estimation of the total error of the distance measurement in the experiment (see Fig. 4).
|
|
DNA Mapping with NALMS Microscopy. Fig. 7 shows a fluorescence image of
-DNA bound by Alexa 488-labeled bis-PNA. Bis-PNA binds tightly to specific recognition sequences in double-stranded DNA (23). The pattern of these binding sites produces a physical map that would be specific for different DNA sequences. NALMS microscopy offers a method for resolving fluorescent probes bound to specific recognition sites in DNA. Under the long-exposure binding conditions used, the 7-base bis-PNA probes bind to more than just their specific recognition sites. Although this occurrence prevents reconstruction of an accurate map, the high number of bound dye molecules provides multiple single fluorophores that overlap in conventional microscopy, and thereby tests the resolving power of NALMS. In Fig. 7, every dye in the overlabeled region (i.e., rightmost part of the image) can be resolved and localized to within 10 nm (most within 7 nm and some within 2 nm), which is far beyond the capability of traditional optical microscopy or previous optical mapping studies. Thus, NALMS offers a resolution method for very-high-density physical mapping of individual DNA molecules.
|
| Conclusion |
|---|
|
|
|---|
2.5-nm precision. The empirically determined limit of the resolution of NALMS microscopy is 8 nm or better, covering the gap between FRET-based (<10 nm) and diffraction-limited microscopy (>100 nm) measurements of the distance between two fluorophores. In the present study, up to five spectrally identical fluorophores could be resolved; however, this is not necessarily the upper limit. In the experiment, we did not observe any influence of the 2-d spatial arrangement of the dye molecules on the resolution. However, for a diffraction-limited spot with more dye molecules, it will be harder to resolve every dye molecule in it, and the precision of the centroid localization for each dye molecule will in general deteriorate. The limit depends on the experimental conditions. The average total number of photons detected before photobleaching for dye molecules varies, depending on the dye molecules and buffer conditions, the N.A., and efficiency of the imaging system. The number of detected photons generally determines the precision of the centroid localization and the limit of the resolution. To reach the maximum resolving power, two conditions would help; lowering the laser illumination power, and using a faster CCD detector. Both conditions increase the probability to catch every photobleaching event. Of course, there are limiting factors, such as the background noise, the CCD noise, the quantum nature of the dye emission, and the sensitivity of the CCD. Also, active control of the drift in the microscope system will greatly help the measurement precision.
The requirement for application of NALMS microscopy is that individual photobleaching events are resolved and that the individual fluorophores are essentially stationary in time and space. Thus, NALMS microscopy as performed here is limited to the measurement of static and dilutely labeled samples. Still, important questions can be addressed. NALMS microscopy will be a powerful approach for the creation of physical maps of DNA. It will be useful in obtaining high-resolution distributions of fluorescently labeled DNA-binding proteins. With the use of CCD detectors with on-chip gain and faster readout, NALMS microscopy will have application to study slow dynamics of molecular machines and even establish spatial distributions of fluorescently tagged macromolecules and assemblies in vivo.
| Acknowledgements |
|---|
| Footnotes |
|---|
Abbreviations: NALMS, nanometer-localized multiple single-molecule; FRET, fluorescence resonance energy transfer; PNA, peptide nucleic acid; TIRF, total internal reflection fluorescence; CCD, charge-coupled device.
Note. While this paper was under review, a paper appeared (27) reporting the use of photobleaching to resolve two fluorophores within the diffraction limit. Our work extends the concept to the nanometer-precision localization of multiple fluorophores that are unresolved in a static image. The image processing used in the present paper differs from that reported in ref. 27. These differences in numerical methods may affect the precision of fluorophore localization in certain circumstances and also the computation time for analyzing the images, particularly when multiple fluorophores are involved. As shown in this work, ultraresolution of multiple fluorophores will be important in many applications, including DNA mapping.
Present address: David Geffen School of Medicine, University of California, Los Angeles, CA 90095. ![]()
|| To whom correspondence should be addressed at: 5735 South Ellis Avenue, SCL 033, Chicago, IL 60637. E-mail: nfschere{at}uchicago.edu.
© 2004 by The National Academy of Sciences of the USA
| References |
|---|
|
|
|---|
This article has been cited by other articles in HighWire Press-hosted journals:
![]() |
A. Agrawal, R. Deo, G. D. Wang, M. D. Wang, and S. Nie Nanometer-scale mapping and single-molecule detection with color-coded nanoparticle probes PNAS, March 4, 2008; 105(9): 3298 - 3303. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Zhang, D. Shu, F. Huang, and P. Guo Instrumentation and metrology for single RNA counting in biological complexes or nanoparticles by a single-molecule dual-view system RNA, October 1, 2007; 13(10): 1793 - 1802. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. R. Selvin, T. Lougheed, M. Tonks Hoffman, H. Park, H. Balci, B. H. Blehm, and E. Toprak Fluorescence Imaging with One-Nanometer Accuracy (FIONA) CSH Protocols, October 1, 2007; 2007(20): pdb.top27 - pdb.top27. [Abstract] [Full Text] |
||||
![]() |
M. Bates, B. Huang, G. T. Dempsey, and X. Zhuang Multicolor Super-Resolution Imaging with Photo-Switchable Fluorescent Probes Science, September 21, 2007; 317(5845): 1749 - 1753. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. E. Moerner Single-molecule Chemistry and Biology Special Feature: New directions in single-molecule imaging and analysis PNAS, July 31, 2007; 104(31): 12596 - 12602. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. W. Hell Far-Field Optical Nanoscopy Science, May 25, 2007; 316(5828): 1153 - 1158. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Sharonov and R. M. Hochstrasser Wide-field subdiffraction imaging by accumulated binding of diffusing probes PNAS, December 12, 2006; 103(50): 18911 - 18916. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Heintzmann and G. Ficz Breaking the resolution limit in light microscopy Brief Funct Genomic Proteomic, December 1, 2006; 5(4): 289 - 301. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. C. Lagerholm, L. Averett, G. E. Weinreb, K. Jacobson, and N. L. Thompson Analysis Method for Measuring Submicroscopic Distances with Blinking Quantum Dots Biophys. J., October 15, 2006; 91(8): 3050 - 3060. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Betzig, G. H. Patterson, R. Sougrat, O. W. Lindwasser, S. Olenych, J. S. Bonifacino, M. W. Davidson, J. Lippincott-Schwartz, and H. F. Hess Imaging Intracellular Fluorescent Proteins at Nanometer Resolution Science, September 15, 2006; 313(5793): 1642 - 1645. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Ram, E. S. Ward, and R. J. Ober Beyond Rayleigh's criterion: A resolution measure with application to single-molecule microscopy PNAS, March 21, 2006; 103(12): 4457 - 4462. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. S. Churchman, H. Flyvbjerg, and J. A. Spudich A Non-Gaussian Distribution Quantifies Distances Measured with Fluorescence Localization Techniques Biophys. J., January 15, 2006; 90(2): 668 - 671. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Balci, T. Ha, H. L. Sweeney, and P. R. Selvin Interhead Distance Measurements in Myosin VI via SHRImP Support a Simplified Hand-Over-Hand Model Biophys. J., July 1, 2005; 89(1): 413 - 417. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. SMITH, T. R. SOSNICK, N. F. SCHERER, and T. PAN Efficient fluorescence labeling of a large RNA through oligonucleotide hybridization RNA, February 1, 2005; 11(2): 234 - 239. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. S. Churchman, Z. Okten, R. S. Rock, J. F. Dawson, and J. A. Spudich Single molecule high-resolution colocalization of Cy3 and Cy5 attached to macromolecules measures intramolecular distances through time PNAS, February 1, 2005; 102(5): 1419 - 1423. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||