Nanoscale force sensing of an ultrafast nonlinear optical response
Edited by Jianming Cao, Florida State University, Tallahassee, FL, and accepted by Editorial Board Member Zachary Fisk July 2, 2020 (received for review March 2, 2020)

Significance
Optical properties of materials are governed by nanoscale charge motion induced by ultrafast light fields. Here, we demonstrate that the electrostatic force originating from light-induced electron motion in a dielectric can be measured using atomic force microscopy. We observe changes in the force originating from second-order nonlinear optical interactions in the sample on a sub-15-nm scale and 100-fs time resolution. The time resolution is set by the light pulse characteristics, not by the properties of the force sensor. Our method does not rely on nonlinear tip–sample interactions, thus advancing the possibility to observe ultrafast nonlinear light–matter interactions with spatial resolution in any material. We envision this technique will allow the correlation of nanometer structure with light-induced time-resolved kinetics.
Abstract
The nonlinear optical response of a material is a sensitive probe of electronic and structural dynamics under strong light fields. The induced microscopic polarizations are usually detected via their far-field light emission, thus limiting spatial resolution. Several powerful near-field techniques circumvent this limitation by employing local nanoscale scatterers; however, their signal strength scales unfavorably as the probe volume decreases. Here, we demonstrate that time-resolved atomic force microscopy is capable of temporally and spatially resolving the microscopic, electrostatic forces arising from a nonlinear optical polarization in an insulating dielectric driven by femtosecond optical fields. The measured forces can be qualitatively explained by a second-order nonlinear interaction in the sample. The force resulting from this nonlinear interaction has frequency components below the mechanical resonance frequency of the cantilever and is thus detectable by regular atomic force microscopy methods. The capability to measure a nonlinear polarization through its electrostatic force is a powerful means to revisit nonlinear optical effects at the nanoscale, without the need for emitted photons or electrons from the surface.
Defects determine the properties of real-world materials. As an example, in optoelectronic materials, structural imperfections lead to charge traps, recombination sites, and other efficiency limitations when converting light to electricity (or vice versa). These defects ultimately determine the lifetime and thus economic viability of the material for a particular application. A major experimental challenge is the characterization of the structure–property relation of the diversity of defects in a given material as a function of processing conditions or composition. On a more fundamental level, the motion of single molecules, conformation changes during chemical reactions, electron dynamics in solids, and the effects of defects or trap states on electron motion and behavior are among many fundamental processes waiting to be observed at the femtosecond and nanometer scale (1). Achieving simultaneous spatial and temporal resolution on the relevant length (nanometer) and time scales (10 fs to 100 fs) is a holy grail in material science. Atomic force microscopy (AFM) can be used to obtain atomically resolved structural data on insulators, semiconductors, and metals in any environment, including in physiological solutions.
Here, we demonstrate measurements on the femtosecond time scale of the optically induced polarization in lithium niobate by force detection using time-resolved AFM (tr-AFM) at room temperature in ultrahigh vacuum (UHV). We take advantage of the second-order nonlinear susceptibility of lithium niobate to generate an electric polarization induced by two laser pulses with ∼100-fs width. The change in electric polarization in the sample is measured as a function of delay times through the change in force by using noncontact AFM (nc-AFM). In AFM, the spatial resolution is determined by the tip radius, in our system, typically on the order of a few nanometers.
Results
We split the output of a mode-locked femtosecond fiber laser operating at 80-MHz repetition rate and 780-nm central wavelength, 200 mW (Toptica FemtoFiber pro NIR), to generate two coherent pulse trains, with a well-defined temporal delay between the two. The pulse trains can be arranged in a noncollinear or collinear fashion and focused onto the tip–sample junction of an AFM incorporated in a UHV system. As the two pulses are delayed with respect to each other, the light intensity at the tip–sample junction oscillates, in delay time, with the period of the laser illumination (
(A) Illustration of the experimental setup and beam alignment. An ultrafast laser is incorporated into an UHV-AFM. The laser beam is split, and one beam is time delayed, before both beams are recombined and focused into the UHV chamber. The beam impinges on the sample surface with an angle of
Lithium niobate is an insulator with a band gap of 3.78 eV (2), a noncentrosymmetric crystal structure, and a large second-order optical nonlinear susceptibility
More generally, when two nonresonant intense electric fields are incident on a material with a nonzero second-order susceptibility, the resultant nonlinear polarization is described by (4)
Second-order interactions between pulses with the same center frequency lead to quasi-DC polarizations. Here we separate these polarizations into two parts: those arising from interactions of a pulse with itself
tr-AFM Measurements on Lithium Niobate.
Lithium niobate as a bulk crystal is extensively used in nonlinear optics and, with its large band gap, is a good benchmark for nonresonant optical nonlinearities.
In frequency modulation AFM (FM-AFM), a cantilever oscillating (6-nm amplitude) at its resonance frequency is brought into close proximity to the sample; tip–sample forces are thus measured by recording the resonance frequency shift of the cantilever (7). The z-cut crystal of lithium niobate, in conjunction with p-polarized incident light, allows for the strongest induced polarization occurring along the surface normal. Fig. 2A shows the recorded frequency shift while the two beams are delayed with respect to each other. An oscillating response (period 2.6 fs) during pulse overlap is measured as expected (
Frequency shift of the FM-AFM measurement with (A) collinear and (C) noncollinear laser configuration; the signal is generated by the polarization change in LiNbO3 while the two laser beams are delayed with respect to each other. A pulse width of
Using an AFM tip as our force detector not only grants us high spatial resolution; in addition, the small probing volume of the AFM tip ensures that phase matching is always satisfied (Fig. 2). As expected, the interferometric optical autocorrelation exhibits fast oscillations (Fig. 2B), while the noncollinear autocorrelation (Fig. 2D) simply traces the intensity overlap of the two pulses. The concurrently recorded AFM measurement, on the other hand, shows a fast oscillating signal in collinear as well as a noncollinear arrangement. While a traditional macroscopic measurement of a noncollinear autocorrelation has no phase sensitivity due to the spatial averaging between the two pulses, the AFM measurement detects the interference of the pulses from a much smaller volume compared to the scale of the optical interference fringes.
The incorporation of the ultrafast laser source does not necessitate modifications on the AFM, and thereby the microscopy retains its intrinsic nanometer spatial resolution. Thus, the tr-AFM signal can be spatially mapped by repeating the pump–probe delay scan at each point on the sample surface. Fig. 3A shows a topography image recorded in FM-AFM mode of an HF-etched periodically poled lithium niobate (PPLN) sample (for large-scale image, see SI Appendix, Fig. S5). The etching creates a ∼120-nm height difference between the regions with different poling directions (8). The height changes gradually over a lateral extent of 200 nm to 400 nm, depending on the location (see height profile in SI Appendix, Fig. S5). The dotted line indicates the location of each delay scan. The amplitude of the tr-AFM signal shows a strong attenuation, which is likely due to a surface adsorbent situated at the boundary between two oppositely polarized regions (Fig. 3C), and is recovered when the upper terrace of the PPLN is reached. The tr-AFM signal is normalized by the optically measured autocorrelation intensity to account for any small fluctuation in laser power during the measurement time. To illustrate the significant drop of the ultrafast signal within a few tens of nanometers, the full delay scans taken at the two positions are plotted in Fig. 3D. Within two consecutive steps of 12.25 nm each (see line scan in Fig. 3C), the tr-AFM signal drops by 30%, followed by another drop of 40%. This clearly demonstrates that the tr-AFM signal can be used to probe spatial variations of the sample response on a sub-15-nm scale in our setup.
Spatially resolved measurement of the ultrafast light-induced AFM signal. (A) The topography is shown, indicating the position of the line scan. At each point (separated by 12.25 nm), a full delay scan is performed. (C) The oscillation amplitude is plotted at each point, with (B) the extracted topography. The vertical lines indicate the extent of the surface adsorbent. (D) Raw data from the tr-AFM scans are shown, corresponding to the red and blue markers in the line scan (A and C), respectively. (E) Constant height nc-AFM measurement with minimal achievable delay. The full oscillation can be clearly resolved with a minimum resolvable step size of about
We note that the fast oscillations observed in the AFM scan can be used to determine the minimal achievable delay step that results in a resolvable frequency shift change. Fig. 3E is a measurement of the cantilever frequency shift at a constant height while the delay between the two pulse trains is swept. Two data points are taken at each delay step, with each averaged for 2 ms. We can distinguish two data points separated by the minimum delay step achievable with our double-wedge time delay setup. The minimal measurable delay time is 25 as due to the delay setup and not the sensitivity or frequency stability of the AFM itself. We thus conclude that time resolution in our experiments is limited by the pulse length of the pump and probe and not by the AFM detector.
Power-, Polarization-, and Distance-Dependent tr-AFM Measurements.
To further characterize the tr-AFM signal and its light-induced origin, various optical power-, polarization-, and tip–sample distance-dependent measurements are performed. First, the amplitude of the signal is recorded at different tip–sample distances to determine how the tr-AFM signal decays along the surface normal. The same measurements were performed using s and p polarization. When p polarization is used (Fig. 4B), the tip–sample force extends over a range of 700 nm, while, with s polarization, it decays toward 0 frequency shift within 200 nm. This can be explained by the surface or bulk response when using s or p polarization, respectively. Lithium niobate exhibits a surface nonlinear response resulting in a p polarization when driven with s polarization (9). This surface response is observed to be weaker in amplitude, as seen with s polarization in our measurement (Fig. 4). In Fig. 4A, the normal FM-AFM frequency shift signal follows the same trend. Both the tr-AFM and the regular nc-AFM distance-dependent spectra are fit to the electrostatic force from Eq. 4 with a tip–sample capacitance according to ref. 10 (see Materials and Methods for details). The distance-dependent results in Fig. 4A show that the photoinduced signal is present at both small and large gap spacings, and thus cannot be due to a thermal expansion effect of the tip or sample. Additional evidence of the electrostatic nature of the signal is given by a measurement on an island of
(A and B) Frequency shift and tr-AFM signal amplitude vs. z-piezo position for two different optical polarizations. S polarization corresponds to the most optimal excitation along the z direction of the PPLN crystal, resulting in a stronger light-induced polarization. All measurements are fitted (solid line) to an electrostatic model. (C) Power-dependent measurement of the tr-AFM signal. A second-order polynomial is fitted to the data according to Eqs. 2 and 3.
The tr-AFM signal was also measured as a function of the input laser power, as shown in Fig. 4C. The observed behavior can be fit by Eqs. 2 and 3. Note that the second-order polarization depends linearly on the intensity of the input beams. This second-order polarization leads to an additional term in the tip–sample potential (Eq. 3), whereby
Discussion
In summary, we show that force measurements can be used to detect the 100-fs laser pulse-induced polarization response in a material. Surprisingly, the time resolution of this force measurement is limited by the laser pulse length, and not by the 300-kHz AFM cantilever resonance frequency. Scanning the AFM tip then allows mapping of the spatial variation of the sample response. In particular, we have observed a large change of the polarization response of a PPLN crystal over a distance of 15 nm. By imaging the response of a few layers of
From our experiments, we conclude that the
Force probing of the nonlinear optical response at the nanometer scale is insensitive to phase matching conditions, making nonlinear optical experiments less challenging. The presented tr-AFM detection of nonlinear optical responses complements emerging scanning probe techniques, such as THz scanning tunneling microscopy (12⇓⇓–15), photoinduced force microscopy (16⇓–18), scanning near-field optical microscopy (14, 19⇓–21), or AFM for surface voltage measurements (22⇓–24), to investigate light–matter interactions on the nanoscale with high temporal resolution.
Material and Methods
AFM Setup.
A commercial JEOL JSPM-4500A UHV system with a cantilever beam deflection AFM configuration is used in this study. A band-pass filter is mounted in front of the four-quadrant photodiode of the deflection system to prevent stray light from the femtosecond laser interfering with the AFM measurement. A Nanonis OC4 system is used for control of the AFM. The femtosecond-pulsed laser is coupled into the UHV chamber via a homebuilt in-vacuum beam steering system, consisting of a mirror on a sphere controlled by stick–slip piezo motors. The laser spot size on the sample surface is approximately 900 μm by 200 μm, with an estimated 2,000 to 9,000 photons per probe area (assumed to be a 10-nm-diameter circle) in 10 ms, depending on the average power used.
Fast Fourier Transform Analysis for tr-AFM Amplitude Extraction.
A fast Fourier transform (FFT) is performed on the frequency shift versus delay time data. From the FFT, the peak amplitude around 384 THz is extracted, which corresponds to the wavelength of the incident laser. The delay scan is performed with the highest resolution of the linear delay stage of 0.6 fs. The raw data are downsampled from 0.6-fs step size to 1.2-fs step size to prevent any false frequency response due to randomly missed/double steps of the delay stage.
Single-Cycle Measurement.
The single-cycle measurement was performed while the z feedback was turned off. The tip was lifted by 0.5 nm from –400-Hz frequency shift set point and held at a constant piezo position while the delay was swept. It was verified that, during the full sweep of the delay, no significant drift of the piezo occurs. The temporal delay between the pulses is controlled by a linear delay stage (Thorlabs ODL220). To achieve delays below 0.6 fs, two quartz wedges are used: One wedge is fixed while the second is moved into the beam path to generate an overall delay in the beam with a minimal step size of a few tens of attoseconds.
Distance-Dependent Measurement.
The distance-dependent measurements were performed by approaching to different frequency shift set points ranging from –222 Hz to –2 Hz. The delay scan is performed with the z feedback on with a very slow feedback loop to just correct for long-term drift as a delay scan over the full pulsed overlap is recorded. The average z position is used for determining the tip–sample distance, with the closest approach set to 0 nm. One should note that this does not reflect the absolute tip–sample distance. The polarization was rotated by 90° (s polarization) using a polarizer and wave plate to keep the power constant at 22.3 mW and 22.8 mW, respectively.
For fitting the electrostatic model, we use (10)
Contribution of Second-Order Polarization to the Measured Frequency Shift.
As mentioned in the Results, Eq. 1, for the second-order nonlinear polarization, results in the expression for multiple nonlinear processes. They can be written as
Due to the quadratic contribution to the force,
Based on Eqs. 1 and 3, we can qualitatively model the response expected from
Spatially Resolved Measurement.
The AFM tip is held at constant frequency shift with a slow z feedback to correct for long-term drift during the delay scan at each point. From each delay scan, the tr-AFM signal amplitude is extracted according to the FFT analysis explained above. The extracted tr-AFM amplitude is normalized by the optical autocorrelation intensity to account for any small fluctuation in laser power (average 7.29 mW) during the measurement time.
AFM Autocorrelation.
In our experiment, a metal-coated silicon tip (Nanosensors, PPP-NCHPt,
Optical Autocorrelation.
A β-barium borate crystal is used for optical autocorrelation measurements through SHG for both intensity (noncollinear) and interferometric (collinear) autocorrelation as shown in Fig. 1. A photodiode is used for detection of the second harmonic signal, with either a spatial or low-pass filter to separate the signal from the fundamental. A Gaussian pulse is fitted to extract the pulse duration. An FFT with a low-pass filter is applied to the interferometric autocorrelation to extract the intensity autocorrelation contribution. Built-in envelope extraction is used in MATLAB to extract the envelope of the AFM autocorrelation signal. The same fitting procedure used for the optical autocorrelation is applied to the extracted envelope to retrieve the pulse width.
Chopper Measurements.
The measurements on MoSe2 are performed with a chopper wheel in one arm of the interferometer. The use of the chopper wheel results in the modulation of the AFM frequency shift at the chopper frequency. This modulation is detected with a Zurich Instruments lock-in amplifier (UHF). The lock-in signal is directly proportional to the effects due to illumination without any artifacts due to drift of the AFM tip during tip lift measurements (25).
Lithium Niobate.
The z cut of the crystal, in conjunction with p-polarized incident light, allows for the strongest induced polarization occurring along the surface normal.
M o S e 2 Sample Fabrication.
The monolayer
Data Availability.
All relevant data supporting the findings of this study are available at Figshare (https://doi.org/10.6084/m9.figshare.12478334).
Acknowledgments
We acknowledge Christopher Phillips for providing the PPLN sample. This work was supported by Natural Sciences and Engineering Research Council of Canada, Fonds de recherche du Québec – Nature et technologies, and Canada Foundation for Innovation.
Footnotes
- ↵1To whom correspondence may be addressed. Email: zenos{at}physics.mcgill.ca or peter.grutter{at}mcgill.ca.
↵2Present address: Department of Physics, ETH Zurich, 8093 Zurich, Switzerland.
Author contributions: Z.S., D.G.C., and P.G. designed research; Z.S., R.R., R.P., A.S., and Y.M. performed research; P.N. contributed new reagents/analytic tools; Z.S. and R.R. analyzed data; Z.S., R.R., R.P., A.S., and Y.M. maintained and improved the experimental setup; P.N. fabricated the MoSe2 sample; P.G. supervised the experiment; and Z.S., R.R., D.G.C., and P.G. wrote the paper.
The authors declare no competing interest.
This article is a PNAS Direct Submission. J.C. is a guest editor invited by the Editorial Board.
This article contains supporting information online at https://www.pnas.org/lookup/suppl/doi:10.1073/pnas.2003945117/-/DCSupplemental.
- Copyright © 2020 the Author(s). Published by PNAS.
This open access article is distributed under Creative Commons Attribution-NonCommercial-NoDerivatives License 4.0 (CC BY-NC-ND).
References
- ↵
- M. Peplow
- ↵
- A. Dhar,
- A. Mansingh
- ↵
- ↵
- R. W. Boyd
- ↵
- ↵
- ↵
- ↵
- N. Argiolas et al.
- ↵
- T. J. Sono et al.
- ↵
- S. Hudlet,
- M. Saint Jean,
- C. Guthmann,
- J. Berger
- ↵
- C. Tam Le et al.
- ↵
- T. L. Cocker et al.
- ↵
- ↵
- M. A. Huber et al.
- ↵
- ↵
- J. Jahng et al.
- ↵
- J. Jahng et al.
- ↵
- B. T. O’Callahan,
- J. Yan,
- F. Menges,
- E. A. Muller,
- M. B. Raschke
- ↵
- ↵
- M. Eisele et al.
- ↵
- T. Jiang,
- V. Kravtsov,
- M. Tokman,
- A. Belyanin,
- M. B. Raschke
- ↵
- ↵
- M. Takihara,
- T. Takahashi,
- T. Ujihara
- ↵
- Z. Schumacher,
- Y. Miyahara,
- A. Spielhofer,
- P. Grutter
- ↵
- Z. Schumacher,
- A. Spielhofer,
- Y. Miyahara,
- P. Grutter